What Are NMES Parameters? The Numbers the FDA Asks Makers to Report

Short answer: the FDA has a guidance document for companies that make powered muscle stimulators. It asks each maker for “a list of all output parameter settings (e.g., amplitude, pulse width, frequency, etc.).” Its output table also lists the waveform, the maximum voltage and current, the charge per pulse, the current density, and the on time and off time.

This page walks through that list in the FDA’s own words. It does not recommend any setting, and it makes no promise about what any device will do. For the basics first, read What is NMES? (NMES stands for neuromuscular electrical stimulation. The FDA document uses the term “powered muscle stimulator.”)

Key facts

  • The FDA’s “Guidance Document for Powered Muscle Stimulator 510(k)s” was issued on June 9, 1999.
  • It is labeled “Guidance for Industry, FDA Reviewers/Staff and Compliance.” It is written for makers and FDA reviewers, not for the people who use the devices.
  • It asks for the output numbers “for each output mode,” and a device can have more than one mode.
  • It says the maximum power density “should be less than 0.25 Watts/cm2 to reduce the risk of thermal burns.”

The FDA document behind the list

The document is called “Guidance Document for Powered Muscle Stimulator 510(k)s.” Its cover says “Document issued on: June 9, 1999.” The FDA’s web page for it says “Content current as of: 03/13/2018.” It covers powered muscle stimulators “identified and classified under 21 CFR 890.5850.”

Because it is written for makers and reviewers, it names each number but does not explain it for users. It also does not say what any setting should be for a person. The sections below follow the document’s own headings.

The output table, item by item

In its section called “Output Specifications,” the guidance gives a table to fill in for each output mode. The items include:

A note under the table says: “For continuously variable parameters, specify the full range; for parameters with discrete settings, specify all available selections.” In plain words, a maker is asked to say which numbers a device can change and by how much.

Output modes

The guidance says: “An output mode is defined (for reporting purposes) as a version of a waveform produced by the unit.” It gives an example: “biphasic symmetrical, biphasic asymmetrical, and monophasic would all be considered separate output modes.”

A device can have more than one mode, and each mode has its own numbers. The guidance asks for the full set of numbers for each one.

Frequency (Hz)

The table lists “Frequency (Hz).” For burst mode, it also asks for “Pulses per burst” and “Bursts per second.”

Pulse width

The table lists “Pulse Width (specify units),” so the maker has to name the unit it uses. For multiphasic waveforms, the table also asks for “Phase Duration (include units)” and to “state range, if applicable.”

Voltage, current and charge

The table lists “Maximum Output Voltage (specify units)” and “Maximum Output Current (specify units).” It asks for each one at three loads: 500 ohms, 2,000 ohms (2k) and 10,000 ohms (10k).

The table also lists “Net Charge” per pulse and “Maximum Phase Charge.”

Waveform and shape

In its section called “Output Waveforms,” the guidance says: “For each output mode, a minimum of four oscilloscope tracings (or accurate diagrams) describing the electrical output waveform should be provided.” Each one should come with “a list of all output parameter settings (e.g., amplitude, pulse width, frequency, etc.).”

The output table gives examples. For the waveform: “pulsed monophasic, biphasic.” For the shape: “rectangular, spike, rectified sinusoidal.”

On time, off time and duty cycle

The table asks for “ON Time (seconds)” and “OFF Time (seconds).” Under burst mode, it lists “Duty Cycle [Line (b) x Line (c)].” In that list, line (b) is “Bursts per second” and line (c) is “Burst duration (seconds).”

The guidance’s section called “Basic Unit Characteristics” also asks for a “Timer Range (minutes).”

Electrodes and current density

The pads, called electrodes, are part of the picture too. In its section called “Description of Accessories,” the guidance says: “The type and size of all available and recommended electrodes should be specified.”

A note under the output table says: “Maximum current density and power density values should be calculated using the conductive surface area of the smallest electrodes provided/recommended for use with the unit.” The same note says: “The maximum power density should be less than 0.25 Watts/cm2 to reduce the risk of thermal burns.”

The guidance’s labeling section adds: “Powered muscle stimulators should be used only with the leads and electrodes recommended for use by the manufacturer.”

Where a device’s own numbers are written down

The guidance’s labeling section, “Attachment I,” lists what a user manual should contain. The list includes “A description of all features, functions, output modalities, and specifications” and “Directions for use.”

The same section suggests this precaution for labeling: “Electrode placement and stimulation settings should be based on the guidance of the prescribing practitioner.”

The FDA’s consumer page on electronic muscle stimulators says: “It is very important that these devices be properly designed, manufactured, and labeled,” and that the person using one “follows the instructions carefully.”

So start with the user manual that comes with your own device. If a number is missing or unclear, ask the maker or your clinician before you use the device.

Safety basics

The FDA’s consumer page says the “FDA has received reports of shocks, burns, bruising, skin irritation, and pain” with some of these devices. It also says: “There have been a few recent reports of interference with implanted devices such as pacemakers and defibrillators.” The page carries the date line “Content current as of: 12/17/2017.”

The guidance says: “Skin irritation and burns beneath the electrodes have been reported with the use of powered muscle stimulators.” Its warnings say stimulation “should not be applied over the neck or mouth,” and should not be applied “transthoracically” (across the chest) or “transcerebrally” (across the head).

Its precautions add: “Portable powered muscle stimulators should not be used while driving, operating machinery, or during any activity in which involuntary muscle contractions may put the user at undue risk of injury.” And: “Powered muscle stimulators should be kept out of the reach of children.”

The FDA’s consumer page says: “Medical device malfunctions can be reported directly to the manufacturer.” It adds: “You can also report to MedWatch, the FDA’s voluntary reporting program.”

Questions to ask your clinician

  1. Which settings on this device can I change, and which stay the same?
  2. What does each number in the manual mean, in plain words?
  3. Where do the pads go, and what size should they be?
  4. What should I feel during a session, and what means I should stop?
  5. Is there anything in my history that means I should not use it, or not on part of my body?

Frequently asked questions

What are NMES parameters?

The FDA’s guidance for makers of powered muscle stimulators asks for “a list of all output parameter settings (e.g., amplitude, pulse width, frequency, etc.).” Its output table also lists the waveform, the maximum voltage and current, the charge per pulse, the current density, and the on time and off time.

Who is the FDA guidance on muscle stimulator settings written for?

It is labeled “Guidance for Industry, FDA Reviewers/Staff and Compliance.” It is written for device makers and FDA reviewers, not for the people who use the devices.

Which NMES setting should I use?

This page does not name one. The FDA’s guidance lists what a maker should report, not what a person should choose. The labeling precaution it suggests says: “Electrode placement and stimulation settings should be based on the guidance of the prescribing practitioner.”

Where can I find the settings for my device?

Start with the user manual that comes with the device. The FDA’s guidance lists “A description of all features, functions, output modalities, and specifications” among the things a user manual should contain. If a number is missing or unclear, ask the maker.

Sources

These are FDA documents. They describe what makers should report and what the FDA has heard about safety. None of them shows what a particular device will do for you.

  1. U.S. Food and Drug Administration. “Guidance Document for Powered Muscle Stimulator 510(k)s” (Guidance for Industry, FDA Reviewers/Staff and Compliance; document issued on June 9, 1999). PDF. Sections used: Introduction; Attachment I, Labeling Guidance (User Manual, Warnings, Precautions, Adverse Reactions); Attachment II, Guidance for Reporting Technological Characteristics (Section 1: Output Waveforms; Section 2: Basic Unit Characteristics; Section 3: Output Specifications and its notes; Section 4: Description of Accessories). https://www.fda.gov/files/medical%20devices/published/Guidance-for-Industry–FDA-Reviewers-Staff-and-Compliance—Guidance-Document-for-Powered-Muscle-Stimulator-510(k)s-(PDF-Version).pdf
  2. U.S. Food and Drug Administration. “Guidance Document for Powered Muscle Stimulator 510(k)s – Guidance for Industry, FDA Reviewers/Staff and Compliance” (guidance web page; issued June 9, 1999; content current as of March 13, 2018). https://www.fda.gov/regulatory-information/search-fda-guidance-documents/guidance-document-powered-muscle-stimulator-510ks-guidance-industry-fda-reviewersstaff-and
  3. U.S. Food and Drug Administration. “Electronic Muscle Stimulators” (consumer information, content current as of December 17, 2017). https://www.fda.gov/medical-devices/consumer-products/electronic-muscle-stimulators

By the NerveOTX editorial team. Last updated October 29, 2026.